Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 52
FREQUENCY
Range
10Hz
100Hz
1000Hz
10kHz
100kHz
1000kHz
10MHz
Sensitivity: sine wave 0.6V RMS (10MHz: 1.5V RMS)
Overload protection: 250V DC or AC RMS

DUTY CYCLE

0.1% ~ 99.9%: ± (2.0%+2) Frequency lower than 10kHz
Sensitivity: sine wave 0.6V RMS
Overload protection: 250V DC or AC RMS
TEMPERATURE (AM-240 ONLY)
Accuracy specification is relative to the user-adjustable temperature offset, and
assumes ambient temperature stable to ± 1°C.
For ambient temperature changes of ± 5°C, rated accuracy applies after 1 hour.
Does not include error of the thermocouple probe
Range
-58ºF ~ +1292ºF
-50ºC ~ +700ºC
Overload protection: 250V DC or AC RMS
Accuracy
± ( 0.1%+5 )
Accuracy
+/-(2.8% + 9ºF)
+/-(2.8% + 7ºC)
6
Resolution
0.01Hz
0.1Hz
1Hz
10Hz
100Hz
1kHz
10kHz
Resolution
1° /1°F

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Am-240

Tabla de contenido