Processing Instructions
Lamineer Tips
WARNINGS
Limitations on processing
INSTRUCTIONS
Initial treatment at the point of use Soak in solution. If debris remains use a cleaning stone suitable for diamond instruments.
Cleaning: Automated
Cleaning: Manual
Disinfection: Automated
Disinfection: Manual
Inspection
Packaging for sterilization
Sterilization
Storage
Use chemicals suitable for diamond coated stainless steel instrument.
See inspection.
Washer disinfector (in accordance with standard ISO 15883).
Without or with mild detergents.
Ultrasonic bath.
Detergent. In accordance with manufacturer's specification for the use of detergent.
Washer disinfector (in accordance with standard ISO 15883).
Without or with mild detergents.
Disinfection bath.
Disinfectant. In accordance with manufacturer's specification for the use of disinfectant.
Thorough rinsing.
Visually inspection for appearance.
Sterile goods packing according to EN 868-5.
Equipment
B-type steam (in accordance with standards EN 13060, EN 285 and EN ISO 17665-1).
Temperature/Time/Pressure
134°C (F 273°F)/Minimum of 3 minutes/3.06 bar (27 psi).
Vacuum drying
Minimum 3 minutes.
Protect from dust, dirt and humidity, UV-light (sunlight) and extreme temperature (above
room temperature).