A 847 For Erlenmeyer Flasks - Miele professional A 612 Instrucciones De Manejo

Módulo inyector
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 49
en - Installation
A 847 for Erlen-
meyer flasks
46
Remove the knurled nut from the screw attachment.
Turn the injector nozzle to insert it into the screw attachment .
Place the support on the frame above the injector nozzle.
Use a 17 mm spanner (WAF 17) to tighten the injector nozzle .
Secure the supports on the right and left using the module's retain-
ing bars .

Publicidad