PRODIGY Voice® Blood Glucose Meter
Caution: The control solution ranges are located
on the back of the Prodigy® test strip vial. They
are not recommended target ranges for your
blood glucose.
Important Information:
• Use only Prodigy® control solutions with your
Prodigy® meter.
• Check the expiration date on the bottle.
Do not use if expired.
• Use within a period of 90 days from the date
that you first open it. Record the discard date
on the control solution bottle when you first
open it to serve as a reminder to discard after
90 days.
• For in vitro diagnostic use.
Why Perform a Control Solution Test:
• To ensure that your meter and test strips are
working properly together.
• To allow you to practice testing without using
your own blood.
When to use:
• Once a week (to ensure that you continue to
have accurate results).