PRODIGY Voice® Blood Glucose Meter
See Figure 22.
Questionable Results:
If the test result falls outside the range printed
on the test strip vial, check the "Trouble-Shooting
Guide" located in the "System Troubleshooting"
section and repeat the test.
Results that fall outside of the expected control
ranges may indicate:
• An error in performing the test
• Control solution bottle not shaken well
• Expired or contaminated control solution
• Control solution that is too warm or too cold
• Test strip deterioration
• Meter malfunction
• Test strips expired
Owner's Manual