Environmental Requirements; Accuracy Specifications; Dc Voltage - Ega Master EGA Tronik 51247 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

- Electro-Magnetic: When carrying out measurement near the electro-magnetic, it may cause
unstable or wrong reading.
- Power: 1 x 9V battery (NEDA1604 or 6F22 or 006P).
- Battery Life: typically 150hours (alkaline battery)
- Sleep Mode (can be disabled)
- Dimensions (H x W x L): 208mm x 76mm x 30mm.
- Weight: Approximate 260g (battery included)
B. Environmental Requirements
- The Meter is suitable for indoor use.
- Altitude: Operating: 2000m
Storage: 10000m
- Safety/Compliances: IEC 61010 CAT.II 600V, CAT.III 300V over voltage and double insulation
standard.
- Temperature and humidity:
Operating: 0ºC~30ºC (≤85%R.H)
Storage: -20ºC~+60ºC (≤85%R.H)
Accuracy:
(a% reading + b digits), guarantee for 1 year.
Operating temperature: 23ºC
Relative humidity: ≤85%R.H
Temperature coefficient: 0.1x(specified accuracy)/1ºC
A. DC Voltage
Range
400.0mV
4.000V
40.00V
400.0V
600V
Remarks: Input impedance: 10M
32
30ºC~40ºC (≤75%R.H)
40ºC~50ºC (≤45%R.H)
ACCURATE SPECIFICATIONS
5ºC
Resolution
0.1mV
1mV
10mV
100mV
1V
Accuracy
Overload protection
(0.8%+3)
(0.8%+1)
(1%+3)
600V DC/AC

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido