3.
Instructions for suction on the patient:
3.1.Connect the accessories, referring to the "CONNECTION DIAGRAM" on the cover.
ATTENTION: in addition to being antibacterial/antiviral, the hydrophobic filter also acts as secondary protection
device and stops any fluids that the primary safety device in the canister lid cannot manage to stop.
Follow the directions below for proper filter assembly:
Side to connect to the VACUUM
inlet (12) of the collection canister
3.2. Switch the device on by following the instructions for the required operation. Always use the device on a flat sur-
face; this will ensure the proper operation of the inflow safety device to prevent the inflow of fluid in the suction
device.
3.3. For more comfortable aspirations, set the desired vacuum value (bar) using the vacuum regulator (4). Turn the
knob clockwise to increase vacuum and turn it anticlockwise to decrease vacuum; these values can be read on
the "vacuum gauge" instrument (3).
3.4. Place your finger on the manual vacuum flow control (15) and, activating it in pulses, begin suction on the pa-
tient using the aspirator tip (NOT SUPPLIED WITH THE DEVICE).
3.5. Once the process is complete, switch off the device.
3.6. Empty and clean the canister and connecting tubes.
Wording:
Aspirator side
Side to connect to the AIR INLET
of the suction machine
14