Minebea Intec Combics Ex Serie Instrucciones De Instalación página 4

Puerto de salida para interfaz
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

English
1
2
4
YDO07-X
Insert the seal in the front panel.
t
y The seal must be seated properly and should not
be cracked.
Re-attach the front panel and secure it with the
t
ten cap nuts with 1 Nm.
Preparing the Cable
t
Strip approx. 14 cm from the end of the cable.
t
Shorten the shielding to approx. 2 cm and pull
back over the insulation.
t
Strip approximately 5 mm of the insulation from
the wires of the connecting cable and affix
ferrules to the wire ends.
Attaching the Cable Entry
Please use extreme caution when performing
any work on the equipment that affects this
5
4
cable gland.
You must use a torque wrench.
The torque for this cable gland is 5 Nm.
t
Remove the protective caps from the bore hole
on the indicator.
t
Insert the included cable gland through the bore
hole and secure from the inside using the locknut
(
t
Insert the cable through the cable gland until the
shielding (
(
gasket (
and cable forms a small beaded rim.
t
Check the shielding and clamps.
3
t
Securely connect the wires of the connecting
cable in accordance with the terminal assign-
ments.
1
).
2
) comes into contact with the clamps
3
). Tighten the screw-down nut (
5
) inserted between the screw-down nut
4
) until the

Publicidad