20. GENERAL SPECIFICATION
Waveform
Amplitude (over 500 Ohm load) 99.5 mA +/- 10%
Max intensity
Output plug
Channels
Max Pulse energy
Batteries
Weight
Dimensions
Safety Classification
Environmental Specifications:
Operating:
Storage:
TYPE BF
APPLIED PART
IP22
Applied Part
Contact duration: At least 10 minutes.
Note: The electrical specifications are nominal and subject to variation from
the listed values due to normal production tolerances of at least 5%.
Asymmetrical rectangular
50V zero to peak across 500Ω
Constant voltage over 470-1500 Ohm
Constant current over 160-470 Ohm
Fully shielded
Single channel
Total output limited to 25 μC per pulse
2 x AA alkaline (two AA batteries) or 2 x AA NiMH
90 g without batteries
102 x 53 x 28 mm
Internal power source.
Temperature range: 0 to +40⁰C
Humidity: 10 to 93% RH non-condensing
Atmospheric pressure: 700hPa to 1060hPa
Temperature range: -10 to +60⁰C
Humidity: 8 to 93% RH non-condensing
Atmospheric pressure: 700hPa to 1060hPa
Equipment providing a degree of protection against
electric shock, with isolated applied part.
Designed for continuous use.
This symbol on the unit means "Refer to Instructions
for Use"
The unit is not water resistant, and should be
protected from liquids.
Complies with EU WEEE regulations
Vaginal and anal electrodes. Optional skin surface
electrode pads. See section 17.