If the totalizer value exceeds the range
99999 the display blinks once per second.
5.4.1Set the input threshold for the totalizer
This value is always represented with
three places after the decimal point. If set
to 0.000 (no threshold) all input values are
processed by the totalizer.
Note: With the ranges 4 ... 20 mA and
2 ... 10 V an interruption at the input signal
may have the undesired effect that the
value in the totalizer goes down;
Menu <–> Selection
Example: 0,100
Select digit
Set digit
press the
key to accept
5.4.2Totalizer Setup
The totalizer adds up the instantaneous
values with a sampling rate of 1 per sec-
ond. These values, when added together,
give a very large resulting number which
normally exceeds the display range.
Hence, it is necessary to apply a conver-
sion to bring the result within the display
range. This is carried out with the help of
factors and scaling.
Explanation: If the display shows a value
of 12.345 and this is totalled over an hour,
then 3.600 values of 12.345 each give a re-
sult of 44.442,000. This value can be dis-
played only if it has been divided by 1.000.
Thus the display shows 44.442.
Page 16
Note: In the event of a power failure the
totalizer value is saved.
this is because the input signal has fallen
below the lowest input signal and this may
be interpreted as a negative instantaneous
value.
The displayed total value will be derived
from the result in the totalizer, leaving the
decimal point free to be set as required.
Note: To optimize the accuracy, it is
recommended to make use of the full
totalizer display range. This is also valid
for the instantaneous value so far as it is
possible to display the total scaling with a
reasonable number of decimal digits (least
possible rounding effects).
The totalizer value can be reset (set to 0)
either via an electrical signal or manually
using the red key.
Example 1:
An empty container is filled at an average
rate of 10 l/s until a volume of 2 m3 is
reached. After this the container is
Instantaneous
Numerical value
value display
at the
totalizer input
X
10 000
Unscaled result
total
in the totalizer
scaling
X
2 000 000
1
Factor x Scaling
Possibilities for setting the
0,1000 x 10
overall scaling
1,0000 x 1
etc.
Note:
In case it is required to output the totalizer
value as weight, the weight per unit volume
(specific gravity) can taken into considera-
tion by setting the factor correspondingly.
emptied and the total value set (reset) to 0.
Instantaneous value
Totalizer value
Number of
Unscaled result
values
in the totalizer
200
2 000 000
Number at the
Totalizer
output of
display
totalizer
2 000
Page 17