When operating close to extreme use or flow rate conditions (close to minimum or
maximum acceptable values), an on-the-spot calibration may be required to suit the real
conditions in which the meter is required to operate.
Calibration factor or "K factor"
Multiplication factor applied by the system
to the electrical pulses received, to transform
these into measured fluid units.
Factory K Factor
Factory-set default factor. It is equal to 1,000.
This calibration factor ensures utmost
precision
in
the
conditions:
Fluid
Temperature
Flow rate
Two procedures are available for changing
the Calibration Factor:
1. In-Field Calibration, performed by means
of a dispensing operation.
2. Direct Calibration, performed by directly
changing the calibration factor.
In calibration mode, the partial and total
dispensed quantities indicated on the display
screen take on different meanings according
to the calibration procedure phase.
In calibration mode, the meter cannot be
used for normal dispensing operations.
In "Calibration" mode, the totals are not
increased.
Samoa Industrial, S.A. · Box 103 Alto Pumarín · 33211 Gijón - Asturias Spain · Tel.: +34 985 381 488 · Fax.: + 34 985 384 163
Calibration
Definitions
following
operating
water / urea solution or liquid food products
20 °C
10 - 30 ltr/min.
Calibration mode
Even after any changes have been made by
the user, the factory k factor can be restored
by means of a simple procedure.
User K Factor
Customized calibration factor, meaning
modified by calibration.
ATTENTION
t
he meter features a non
that keeps the data concerninG calibration
and total dispensed Quantity stored for an
,
indefinite time
even in the case of a lonG
;
power break
after chanGinG the batteries
calibration does not need to be repeated
-
volatile memory
,
.
9
R. 10/12 836 828