Verifying The Gauge Calibration; Batching (Model T) - Elcometer 130 Guia Del Usuario

Medidor de contaminación salina
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
R

6 VERIFYING THE GAUGE CALIBRATION

The Elcometer 130 is factory calibrated. Users can verify the gauge's
performance in the field using the optional Calibration Verification Tiles, part
number T13023980.
Supplied as a set of 3, these tiles can be used to verify the accuracy of the
gauge calibration at 0.4, 5 and 20μg/cm² (nominal values).
To verify the calibration, press Menu/Calibration/Calibration Verification and
follow the on screen instructions.
Should the gauge / tile readings be outside the stated accuracy of the gauge,
re-calibration is recommended - contact Elcometer or your local supplier for
further information.

7 BATCHING (MODEL T)

1
To use the Batching memory facility, press the Batch softkey.
2
Select "New Batch" or "Open Existing Batch" to add readings.
3
Copy and review batch data.
4
Select "Edit Batch" to rename, clear readings from or delete a batch.
5
Fixed Batch Size allows users to pre-define the number of readings to be stored in a batch.
Once all readings have been taken the gauge automatically opens a new batch with a link
to the original batch name. For Example NewBatch_1 becomes NewBatch_2, NewBatch_3,
etc.
11
www.elcometer.com

Publicidad

Tabla de contenido
loading

Tabla de contenido