NOTE:
1
The SpO
upper and lower alarm limits of SpO
2
The PR Hi Alarm and PR Lo Alarm stand for the upper
and lower alarm limits of PR respectively.
3
If the user changes the default value of Lo Alarm or Hi
Alarm, after restarting the oximeter, the value will
resume to the default value for corresponding patient
type.
3.3.1 System Mode
There are two items for selecting:
Waveform mode
Large numeric mode
Then confirm the selection by pressing the Function button.
3.3.2 Model
It indicates that the oximeter can be used by veterinary only and
this item is unselectable and unmodifiable.
3.3.3 Alarm Volume
The Alarm Volume button is used to adjust alarm volume and its
range is from one to five.
When Alarm System is setup to ON, if a low alarm, a medium
alarm or a high alarm occurs, the oximeter sounds beep.
3.3.4 Pulse Volume
The user can turn on or off the pulse volume by pressing Pulse
Volume, and change volume level to 1, 2, 3, 4, 5 or OFF. Press
the Function button to enter setup state, then use the Up Arrow
or Down Arrow button to choose, then confirm it by pressing the
Function button.
Hi Alarm and SpO
2
Lo Alarm stand for the
2
respectively;
2