A2.2 Electromagnetic Immunity - For All Equipment And Systems - Henry Schein Nellcor 900-8273 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

A2.2 Electromagnetic Immunity - for all
EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic
immunity
The oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the
oximeter should assure that it is used in such an environment.
Emissions test
Electrostatic
discharge(ESD)
IEC/EN61000-4-2
Electrical Fast
Transient/Burst
IEC/EN61000-4-4
Surge
IEC/EN61000-4-5
Compliance
±6kV contact
±8kV air
±2kV for
power supply
lines
±1kV for
input/output
lines (>3m)
line to line
line to ground
58
used for domestic
purpose.
Complianc
Electromagneti
e level
c environment -
Floors should be
wood, concrete
or ceramic tile. If
floors are
±6kV
covered with
contact
synthetic
±8kV air
material, the
relative humidity
should be at least
30%.
Mains power
quality
should be that of
N/A
a typical
commercial or
hospital
environment.
N/A
guidance

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hs-vet

Tabla de contenido