Software Updates
From time to time, software updates may be made available to improve operating
performance or add additional features. Some updates will be provided at no charge to
implement operating efficiencies while others will be optional, paid upgrades, to add new
refrigerants etc. Many of the updates can be completed by the user; however some will
require the instrument to be returned to the factory for new gas calibrations.
The Ultima ID Pro™ has a USB update port located on the Back Panel Connections. This
port should not be used for any other purpose other than to install factory updates using
the Mastercool factory USB drive. IF YOU DO NOT REGISTER THE ANALYZER WE
WILL NOT BE ABLE TO INFORM YOU OF ANY SOFTWARE UPDATES!
3.6

Low Battery Warning

A battery warning will occur when the internal battery voltage becomes low (Figure 31). You
can either continue working or connect the AC Power Supply and continue to work by
pressing 'IGNORE. The analyzer will charge the battery while it is connected to the AC
Power Supply but if you wish to use it wirelessly press 'OFF', plug the analyzer into the AC
Power Supply via the 12DC Outlet Input on the Back Panel and allow it to fully charge. Full
charge will take at least 2 hours.
3.7

Air Sensor Low

In the event you receive an 'Air Sensor Low' message like (Figure 32) there is no need to
discontinue use. This message is meant to alert you that the built in oxygen sensor is
depleting and will need to be replaced in the near future. If the analyzer you are using is new
and you believe you received this message in fault, first verify gas is not flowing into the
analyzer and that you are in a well ventilated area. Once you have verified both, press
'RECAL' to recalibrate the analyzer. If you have had the analyzer for a few years the oxygen
sensor may be near depletion. Pressing 'TEST' allows you to bypass the message and finish
your testing. The oxygen sensor is a consumable part and will eventually need to be
replaced. After receiving this message plan on contacting Mastercool Inc to schedule a
service appointment.
Manual Part Number:
5-06-7000-70-2
THE BATTERY
VOLTAGE IS LOW
RECHARGE THE
BATTERY SOON OR
USE 12V POWER
SUPPLY
OFF
IGNORE
Figure 31
AIR SENSOR LOW
DISCONNECT HOSE
FROM TANK
AND PRESS RECAL
RECAL
Figure 32