The technical specifications may change without notice
Shelf life: maximum 5 years from the date of manufacture

ACCESSORIES SUPPLIES

• Complete aspiration jar 4000ml
• Conical fitting
• Transparent tubes set 6 mm x 10mm
• Hydrophobic and antibacterial filter
The filter is made of hydrophobic material and blocks the passage of liquids that come into contact with it.
Replace it whenever you suspect that it may be contaminated and/or it becomes wet or discoloured.
Replace the filter every time it is used if the suction pump is used on patients in unknown pathological situa-
tions and where an assessment of indirect contamination is not possible.
If, however, the patient's pathology is known and/or there is no risk of indirect contamination, the filter should
be replaced after every work shift or once a month even if the device is not used.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine
should comply with ISO 10993-1 standards on material biocompatibility.
WARNING: The medical device is provided without a specific suction probe. If this device must be used with a
specific suction probe, the end user is responsible for making sure it complies with the EN 10079-2 regulation.

CLEANING OF ACCESSORIES

To clean the plastic housing of the device wear disposable latex gloves and clean with denaturated alcohol
or hypoclorite solutions.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
• Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminat-
ing substances;
• Disconnect the jar from the device
• Disconnect all tubes from the jar and the protection filter
• Empty and dispose of the Jar according to the laws in force in your country;
• Separate all parts of the cover (overflow valve, o-ring);
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly.
Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abra-
sive brush to remove incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive).
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization
stem cycle at 121°C (1 bar relative pressure) making sure that the jar is positioned upsidedown.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged.
Assemble the jar as follows:
• Place the overflow valve into its seat in the cover (under VACUUM connector)
• Insert floating valve keeping the o-ring towards the opening of the cage
• Place the o-ring into its seat around the cover
• After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leack-
ages or liquid exit
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C. The conical connector
can be sterilized on autoclave using a sterilization cycle at 121°C.
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
Aspiration jar: the mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and
sterilization. Beyond this limit, the physical-chemical characteristics of the plastic material may show signs of
decay. Therefore, we recommend that you change it.
ENGLISH
8