CLEANING OF THE APPLIANCE
The cleaning of the device shall be carried out with a soft and dry cloth and with non-abrasive deter-
gents. Do not use excessively wet cloths as the contact of liquids with the electrical parts of the appli-
ance may cause malfunctions or may be hazardous.
While cleaning the device, make sure that no liquids get inside the appliance and that the
power outlet is disconnected.
Wait for the device to dry off before storing or using it again.
MAINTENANCE
The Eolo device does not have any part that requires maintenance and/or lubrication.
Nonetheless, it is necessary to carry out some checks to verify the function and security of the appli-
ance before each use. Remove the appliance from the box and always check that there is no visible
damage; particular attention should be given to any cracks in plastics which may leave some electrical
components exposed. Check also the integrity of the power cable which might have been damaged
during the previous use.
Then connect the cable to the power network and switch it on. Close the compression cap with one
finger and check that there are no disturbing noises which may be evidence of malfunction.
Check that the nebulizer has not suffered any breakage during the previous use (it was stored inap-
propriately or has suffered impacts). The appliance is protected by a protection fuse (F 1.6A L 250V)
not accessible from outside. Therefore, contact the technical staff authorized by the manufacturer for
its replacement.
Type of defect
1. Poor atomization.
2. Poor atomization.
3. Lack of
atomization.
4. Slow atomization.
5. Noisy device.
Defects 1 - 2 - 3 -
4 - 5
If the appliance still does not nebulize after having checked the above mentioned conditions, it is rec-
ommended to contact the reseller or Gima's technical service.
BEFORE CARRYING OUT ANY CHECKS IN THE EVENT OF ANOMALIES OR MALFUNC-
TIONS, CONTACT GIMA'S TECHNICAL SERVICE. THE MANUFACTURER OFFERS NO GUAR-
ANTEE FOR THE APPLIANCES WHICH HAVE ASSESSED AS TAMPERED AFTER THE CHECKS
CARRIED OUT BY THE TECHNICAL SERVICE.
CLEANING AND WASHING OF ACCESSORIES
Before each use and / or after the cleaning operations, check the integrity of all accessories supplied
with the device. Turn off the device before each cleaning operation and disconnect the power cable
from the socket.
PREPARATION
1. Pull out the air hose from the nebulizer leaving it plugged to the air outlet pipe of the device.
2. Rotate the upper part of the nebulizer anticlockwise.
3. Disconnect the internal pisper at the bottom of the nebulizer with finger force.
Cause
Clogged ampoule.
Clogged ampoule.
Nozzle stuck.
Excessive oily drug.
Prolonged use.
None of the corrective
actions has been effective
ENGLISH
Corrective action
Clean and disinfect the ampoule as set out in
the manual.
If the washing had no positive outcome, replace
the ampoule.
Press hard the nozzle (cylindrical immersion
tube) placed inside the polycarbonate ampoule
bottom with a finger.
Dilute the drug with saline.
Contact the reseller or Gima's technical service.
Contact the reseller or Gima's technical service.