Operation

6
Operation
On/Off key for switching on respectively switching off the instrument
Scroll key for selection of °F dew point, % relative humidity, °F temperature
Max/Min key for indication of the maximum and minimum values since the point of time
when the instrument was switched on
7

Measurements

7.1

Measurement with measuring chamber, connection via plug nipple

1. Preparation of the measuring point
Let compressed air flow off at the sampling point before measurement in order to remove condensate and
particles. This avoids a soiling of DP 207 and the measuring chamber.
Stagnant air leads to long adjustment times.
If condensate occurs at the measuring point please check the compressed-air conditioning before measure-
ment.
2. Switch on DP 207 and wait until the initialization has been finished.
Please observe the chapter "Operation".
3. Connect the measuring chamber screwed onto DP 207 with the plug nipple coupling of the measuring
point
4. Wait until the value in the display of DP 207 has stabilized. Depending on the position of the measuring
point this may take up to 15 minutes.
5. Disconnect the measuring chamber from the plug nipple coupling of the measuring point after measure-
ment. Switch off DP 207 if you do not want to carry out further measurements.
7.2
Measurement without measuring chamber, connection via external thread G½''
1. Preparation of the measuring point
Make sure that the measuring point is depressurized.
10
DP 207