Invacare IRC1135 Instrucciones De Ensamblado, Instalación Y Operación página 3

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

OPERATION (FIGURE 2)
WARNING
\
Proper operation of this product is necessary
to avoid physical harm. Improper use can
cause personal injury and/or tissue damage.
This product should not be used unless the
operator has been instructed by a physician
or other qualified health care professional.
1 . F a m i l i a r i z e y o u r s e l f w i t h t h e A s p i r a t o r, FIGURE 2.
2 . Attach the suction tubing to the Suction Inlet.
3 . Check to see that the ON/OFF Switch is in the
OFF position.
4 . Turn the Control Knob fully counterclockwise.
5 . Plug power cord into electrical outlet.
6 . Turn the ON/OFF Switch to the ON position.
7 . Turn the Control Knob clockwise to adjust the
Vacuum Gauge to the setting specified by your
physician or qualified health care professional.
NOTE: As the knob is turned counterclockwise the
vacuum will decrease, as the knob is turned clock-
wise the vacuum will increase.
CAUTION
The Collection Jar should be emptied after
every use. NEVER let the fluid level exceed
the "Safe Full" line. Overflow can damage
the Aspirator.
8 . After each use, empty the Collection Jar. R e f e r t o
CARE AND CLEANING i n t h i s i n s t r u c t i o n s h e e t .
Control Knob
ON/OFF
Switch
FIGURE 2 - OPERATION
Vacuum Gauge
Suction
I n l e t
Collection
Jar
3 3 3 3 3
REPLACING THE HYDROPHOBIC
FILTER TUBING ASSEMBLY
(FIGURE 3)
NOTE: Either of the following conditions may re-
quire replacement of the Hydrophobic Filter Tub-
ing assembly: a) the filter is dirty and the aspirator
has lost suction. b) there is liquid in the hydropho-
bic filter tubing assembly due to extended use or
overflow.
1 . Remove the Elbow Connector end of the Hy-
drophobic Filter Tubing assembly from the top
of the Collection Jar.
2 . Unthread the Hydrophobic Filter Tubing assem-
bly from the Vacuum Gauge and discard.
CAUTION
Use only a hydrophobic bacteria filter. DO NOT
replace with other types of bacteria filters.
3 . Install new hydrophobic filter tubing assembly
by reversing STEPS 1-2.
NOTE: Hand tighten the N E W
tubing assembly into the Vacuum Gauge. A loose
filter may decrease suction.
NOTE: If necessary, twist the elbow connector to
align with the collection jar.
Hydrophobic Filter
Tubing Assembly
Vacuum
Gauge
FIGURE 3 - REPLACING THE
HYDROPHOBIC FILTER TUBING
ASSEMBLY
N E W
N E W
N E W hydrophobic filter
N E W
Elbow
Connector
Collection
Jar

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Irc1136

Tabla de contenido