ConT
iMPOrtANt: This will NOt interrupt the shaking function. Press the
on/off button below the speed display to interrupt the shaking function.
Restart the timer by pressing the on/off button below the time display.
Unit will continue counting down to zero (0:00). When the display reaches
zero (0:00), you will hear the four (4) audible beeps that indicate the
count down function is complete and the shaking function will stop.
5. turning unit off:
a. To turn the unit off, press the standby button. The speed and time displays
will be blank, the standby indicator light will illuminate. The stirrer should be
kept in standby mode when not in use. To completely cut off power to the
unit, disconnect the power cord from the unit or un-plug from the wall outlet.

OPErAtiNg tiPS

Built-in memory maintains the last used speed and time settings during a power
interruption.
The stirrer increases speed at a steady rate until the set-point is reached, if
the spin bar is too large or the liquid is too viscous, the stirrer may not reach its
set-point or become de-coupled from the units drive magnet. The set-point speed
needs to be reduced. The unit is designed to ramp slowly to assure that the spin
bar stays coupled.
* The model 2000/200L unit contains a larger spin bar with a center
pivoting ring. Ring may need to be replaced after eight (8) hours of use.
Contact your VWR representative for replacement spin bars or rings.