HEALTHCARE
6.3
SL500-BAT & SL510 Only
Item
Input power requirements to
universal power supply
Output power of universal
power supply
Cordset connector type
Cordset length
Battery voltage and capacity
NIBP measurements per
battery charge
Battery charge time
Battery charge cycles
Physical dimensions
Weight
6.4
SL510 Only
Single temperature
measurement time
Continuous temperature
measurement settling time
Single temperature efficacy
Continuous temperature
accuracy
Temperature probe sensor type
Single temperature
measurement range
Continuous temperature
measurement range
Specification
100 - 240 VAC, 390 mA, 50 - 60 Hz
10 V, 1.25 A
IEC 320 C7 connector
2 m, 6.7 feet
8 V, 3.0 Ah
200 at 2 minute cycle time
12 hours, trickle charge
Height: 13 cm Width: 23 cm Depth: 17 cm
4 kg, including accessories
15 seconds
3 - 4 minutes
Average error within +/- 0.1°C (0.2°F)
Standard deviation within 0.4°C (0.7°F)
Meets ASTM E 1112-00 accuracy
requirements
Temperature Range
Less than 35.8 °C (96.4 °F)
35.8 °C (96.4 °F) and less
than 37.0 °C (98.0 °F)
37.0 °C (98.0 °F) and less
than 39.0 °C (102.0 °F)
39.0 °C (102.0 °F) and less
than 41.0 °C (106.0 °F)
Greater than 41.0 °C
(106.0 °F)
Thermistor
35°C to 39°C (95.0 to 102.2°F)
28°C to 43°C (82.4°F to 109.4°F)
500
Max. Error
0.3 °C (0.5 °F)
0.2 °C (0.3 °F)
0.1 °C (0.2 °F)
0.2 °C (0.3 °F)
0.3 °C (0.5 °F)
21