Using Your Oc Series Oxygen Concentrator - GCE OC Serie Manual Del Usuario

Ocultar thumbs Ver también para OC Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 59
TRANSPORT STORAGE AND OPERATING CONDITIONS
TEMPERATURE
RELATIVE HUMIDITY
ALTITUDE
ATMOSPHERIC PRESSURE
ENVIRONMENT
2 . USING YOUR OXYGEN CONCENTRATOR
WARNING: Do not use extension cords or electrical adapters.
1 .
Choose a location for your concentrator that allows it to draw in air freely. Make sure
the device is at least 30 cm away from walls, furniture, curtains, or any other item that
may restrict airflow to the device.
2 .
Install the concentrator's two stage filter:
a .
Remove the filter access panel
b .
Insert the inlet filter and turn clockwise to connect
c .
Replace the filter access panel
3 .
Set up the humidifier bottle:
a .
Press the humidifier connector button to remove the humidifier bottle
b .
Remove the cap by turning the humidifier bottle clockwise
c .
Fill the humidifier bottle with distilled or cooled boiled water between the min and
max markings
d .
Replace the humidifier cap and tighten until secure
e .
Plug the humidifier bottle into the humidifier connector to reconnect it to the oxy-
gen concentrator
NOTE: Replace the water in your humidifier bottle on a daily basis .
4 .
After reading and understanding the contents of this manual, plug the power cord into
a grounded wall outlet and press the power switch to the On [I] position.
5 .
Make sure that the humidifier bottle is secure:
a .
Use your finger to gently block the oxygen outlet on the humidifier bottle for 20
seconds
b .
Once the float in the flow meter drops to the bottom of the gauge, remove your
finger.
c .
If the bottle makes a whistling sound - it means that the humidifier is properly se-
cured to your device.
d .
If you cannot hear this sound, remove the humidifier bottle, unscrew the cap, re-
place tightly and repeat the above test.
e .
If no whistling sound is heard after taking these steps, contact your home care
provider.
6 .
Adjust the oxygen flow so that the float mark is centered on the line marking the flow
rate prescribed by your physician:
a .
Turn the flow control dial clockwise to increase the output flow rate.
b .
Turn the flow control dial anti-clockwise to decrease the output flow rate.
7 .
Connect your oxygen tube to the oxygen outlet and put on your cannula as directed
by your home care provider. If you are using an extension tube, connect this to the
oxygen outlet and oxygen tube.
OPERATING
10°C to 37°C (50°F to 98°F) -30°C to -70°C (-22°F to -94°F)
20% to 65%,
no condensation
0 to 1828 meters
80 kPa ~ 101 kPa
Dry, well-ventilated,
dust-free
and pollution-free
Away from electromag-
netic interference
7/167
TRANSPORT
AND STORAGE
15% to 95%,
no condensation
-
80 kPa ~ 101 kPa
Placed upright and vertical
at all times
EN

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Oc-e80Oc-e100

Tabla de contenido