Chapter 8 Troubleshooting Guide; Faqs - KPS TherCam384 Manual De Usuario

Ocultar thumbs Ver también para TherCam384:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Troubleshooting Guide

FAQs

Problems
Failure to start
Big deviation between
indicating value and actual
value of battery capacity
Unclear infrared image
Unclear visible image
Incorrect temperature
measurement
86
User Manual
Causes
Insufficient battery capacity
Poor battery contact
Plug of external power supply is not
inserted correctly
The battery capacity is exhausted
The battery life is over
Focusing is not conducted
The infrared lens has steam or is
polluted.
Too dark environment
The visible lens has steam or is
polluted
The objective is not focused
Incorrect parameter settings related to
temperature measurement
Heterogeneity correction is not
conducted for a long time.
Measure temperature immediately
after starting
No calibration for a long time
Action
Recharge and then use the battery
Take out the battery, put it into the battery
cabin again and install in place
Pull out the power plug, insert again and
push in place
Replace with a battery full charged
Replace with a new battery
Focus manually or automatically to make
the image clear
Use a special tool to clean the lens
Take lighting measures properly
Use a special tool to clean it
Focus manually or automatically to
make the image clear and then read the
temperature.
Change the parameter settings or directly
recover the default parameter
Set the customized button in the menu
as "Compensation", press the actual
customized button. Heterogeneity
correction is conducted when hearing
shutter sound
To ensure the precision of temperature
measurement, we suggest that you wait
for 5-10min before temperature measuring
To obtain precise temperature
measurement results, we suggest that you
returnthe thermal imager for calibration
once a year