Braun VitalScan 1 Manual Del Usuario página 23

Ocultar thumbs Ver también para VitalScan 1:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

All manuals and user guides at all-guides.com
Guidance and manufacturer's declaration – electromagnetic emissions
The ME equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the ME equipment should assure that
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Non-Life Support Equipment Separation Distance Calculation
Rated maximum output
power of transmitter (W)
0.01
0.1
1
10
100
it is used in such an environment.
Compliance
The ME equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
Group 1
Class B
Not Applicable
Not Applicable
(3Vrms / 3V/m compliance)
150 kHz to 80 MHz
in ISM bands
3,5
d = [
] P
V
1
0.12
0.37
1.17
3.69
11.67
Electromagnetic environment – guidance
are not likely to cause any
interference in nearby electronic equipment.
Complies
The ME equipment is
solely battery powered.
Separation distance according to
frequency of transmitter (m)
80 MHz to
800 MHz
3,5
d = [
] P
E
1
0.12
0.37
1.17
3.69
11.67
800 MHz to
2.5 GHz
7
d = [
] P
E
1
0.23
0.74
2.33
7.38
23.33
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Vitalscan 3Bpw200laBpw220la

Tabla de contenido