Guidance and Manufacturer´s Declaration –
Electromagnetic Emission: for all equipment and systems
The Infrared Thermometer IRT3W is intended for use in the electromagnetic
environment specified below. The customer or the user of the Infrared
Thermometer IRT3W should assure that it is used in such an environment.
Emissions
Compliance
test
RF
Group 1
emissions
CISPR 11
RF
Class B
emissions
CISPR 11
Guidance and Manufacturer´s Declaration –
Electromagnetic Immunity: for all equipment and systems
The Infrared Thermometer IRT3W is intended for use in the electromagnetic
environment specified below. The customer or the user of the Infrared
Thermometer IRT3W should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Electrostatic
±6 kV
discharge
contact
(ESD) IEC
±8 kV air
61000-4-2
Power
3 A/m
frequency
(50/60 Hz)
magnetic
field
Electromagnetic environment - guidance
The Infrared Thermometer IRT3W uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The Infrared Thermometer IRT3W is suitable
for use in all establishments, including domestic
establishments and those directly connected to
the public low- voltage power supply network that
supplies buildings used for domestic purposes.
Compliance
Electromagnetic environment-
level
guidance
±6 kV contact
Floors should be wood,
concrete or ceramic tile. If floors
±8 kV air
are covered with synthetic
material, the relative humidity
should be at least 30 %.
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
29