20. ELECTROMAGNETIC DECLARATION
Icare HOME is class B equipment and needs special precautions regarding
EMC and needs to be installed and put into service according to EMC
information provided below.
Guidance and manufacturer's declaration–Electromagnetic emissions
Icare HOME (TA022) is intended for use and should be used in the
electromagnetic environment specified below.
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
flickering emissions
IEC 61000-3-3
Guidance and manufacturer's declaration–Electromagnetic immunity
Icare HOME (TA022) is intended for use and should be used in the
electromagnetic environment specified below.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
Transients/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruption
and voltage
variations on
power supply
lines
IEC 61000-4-11
Group 1
Class B
NOT
APPLICABLE
NOT
APPLICABLE
IEC 60601
Test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±2 kV 100 kHz
repetition
frequency
±1 kV for line(s)
to line(s)
±2 kV for line(s)
to earth
0 % UT for 0.5 cycle
(1 phase)
0 % UT for 1 cycle
70 % UT for 25/30
cycles (50/60 Hz)
0 % UT for 250/300
cycles (50/60 Hz)
Icare HOME (TA022) is battery
operated and uses RF energy only
for its internal function. Therefore,
its RF emissions are low and are not
likely to cause any interference in
nearby equipment.
Icare HOME (TA022) is suitable for
use in all establishments, including
domestic establishments and
those directly connected to public
low-voltage power supply network
that supplies buildings used for
domestic purposes.
Power level of Icare HOME (TA022)
is below standard requirement,
batteries are not rechargeable.
Icare HOME (TA022) batteries are
not rechargeable.
Compliance
Electromagnetic
level
environment-
Guidance
±8 kV contact
Floors should be wood,
±15 kV air
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
NOT
Icare HOME (TA022)
APPLICABLE
tonometer is not
operational when
connected to an
external computer,
Icare HOME (TA022)
batteries are not
rechargeable.
NOT
Icare HOME (TA022)
APPLICABLE
tonometer is not
operational when
connected to an
external computer,
Icare HOME (TA022)
batteries are not
rechargeable.
NOT
Icare HOME (TA022)
APPLICABLE
tonometer is not
operational when
connected to an
external computer,
Icare HOME (TA022)
batteries are not
rechargeable.
11