7. 2 REPLACING THE HANDPIECE SUPPORT

In case of accidental breakage (fall, etc.), the
device is designed to permit the replacement of
the handpiece support
(ref: E 58365).
To replace the handpiece supportlay the control
unit on the opposite side to the support.
A screwdriver of the Pozidrive type is
recommended for this operation.
Proceed as follows:
- Switch off the water supply, switch off the
device (OI position) then unplug the mains plug
(Figure 4, diag. 1).
- On the lower panel of the S
control unit, unscrew the fixing screw (Figure
5, ref. 1) of the handpiece support (Figure 5,
ref 2).
- Release the support from the centring lug
(Figure 5, ref. 3) and the two support lugs on
the upper part of the unit.
- Slide the support along the cord (Figure 5, ref
4) then release it, noting its direction.
- Insert the handpiece cord into the new support
in the direction you noted.
- Slide the support along the cord.
- Put the support in place on the base of the
device, positioning the centring lug and the
supporting lugs on the upper part of the unit
correctly, and aligning the holes for the fixing
screws.
- Insert the screw and tighten (be careful not to
over-tighten).
- Clean the device.
14

7. 3 TROUBLESHOOTING

Please refer to the table on the following page
N
®
UPRASSON
EWTRON