Cleaning and Disinfection

5.1

Safety Instructions

On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 29 psi (2 bar).
Maximum water pressure during rinsing: 29 psi (2 bar).
Do not exceed a temperature of max. 167 °F (75 °C)!
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
Do not sterilize!
5.2

Reprocessing Limitation

Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
5.3

Required Aids

Reprocessing step Items Needed
Precleaning
Soft plastic brush/soft cloth
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(sterile, lint-free)
Compressed air gun (2 bar maximum)