1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Ensure that the golden contact surfaces on the plug are not
visibly contaminated.
3. Move the movable parts back and forth (at least 5 times).
4. Mount the irrigation tube on the shaft tube.
5. Brush the inside of the shaft tube with a round brush (at
least 5 times).
6. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
7.6
Manual cleaning and disinfection
Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 12.3 (handles: max. 11!).
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Prepare a cleaning bath and a separate disinfectant bath.