5.5

Precleaning

Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Screw the Luer-Lock adapter (=irrigation adapter) onto the
argon gas connection of the adapter.
3. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
4. Remove the syringe. Leave the irrigation adapter on the
product for the next processing steps.
5.6

Manual cleaning and disinfection

Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Prepare a cleaning bath and a separate disinfectant bath.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Completely fill a disposable syringe with cleaning solution
and rinse the lumina (at least 5 times).