Oxygen Analyzer
Recommended separation distances between portable and mobile RF communications equipment
X
1
The MiniO
®
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the MiniO
distance between portable and mobile RF communications equipment (transmitters) and the MiniO
ed below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
All manuals and user guides at all-guides.com
and the MiniOX
X
1
®
can help prevent electromagnetic interference by maintaining a minimum
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 8.8 SP
0.88
2.8
8.8
28
88
P/N 711462 (Rev.3) 02/2016
1
®
(m)
80 MHz to 800 MHz
d = 1.2 SP
0.12
0.38
1.2
3.8
12
Section 3, Specifications
X
1
®
as recommend-
800 MHz to 2.5 GHz
d = 2.3 SP
0.23
0.73
2.3
7.3
23
3-4