Operation
WARNING
Danger due to incorrect measurement
Incorrect measured values are displayed if an incorrect
response factor is entered. Reliable measurement is not
ensured.
► Contact Dräger Service when determining the correct
response factor.
NOTICE
Due to the physical influences of the ambient conditions on
the behaviour of oil dissolved in air, deviations other than the
accuracy of the system in general (with regard to isobutylene)
may occur depending on the oil type and the measured value
reading. This was taken into account when determining the
correction factors so that the PID 5000 emits an alarm before
exceeding the limit value. This can mean that an alarm
sounds before a limit value is actually reached.
The approved oil types are listed in the table below.
Designation
Shell Corena S2P150
Shell Corena S2P100
Boge 3000
Other oil types upon request.
22
Change response factor
1. Select the instrument menu (see chapter 4.3 on page 21).
2. Use the magnetic pen to tap on "menu" until "SETTINGS"
appears on the display.
3. Select "change".
4. Select "Response". This accesses the screen for setting
the response factor.
5. Adjust the response factor (1) by
tapping on "more" or "less" and
confirm with "accept".
6. The response factor currently set is
shown on the bottom right of the
display (2).
4.3.2

Alarm

WARNING
Danger due to contaminated air
The delay before contamination is detected is set depending
on the ratio between the volume of the gas line and the
Air Guard or PID 5000 and the duration of the actual
measurement.
This is primarily due to the fact that the volume flow within the
Air Guard has to be greatly reduced for correct measurement,
whereas this does not result in a reduction in the volume flow
to the pressure vessels to be filled.