Use
|
● Measurement module
All settings can be altered at any time via the menu.
4.2

Menu

To access the menu:
● Press the
To access the context menu:
● Press the
To navigate to the menu or to configure the settings:
1. Press the
2. Press the
To exit the menu:
● In the sub-menus, select Back.
● In the main menu, select Measuring.
4.3

Prerequisites for use

The device is adjusted in the factory before delivery. Dräger
recommends having the device tested for accuracy every 12 months
by DrägerService or by persons authorized by Dräger.
● Wait at least 15 minutes after consuming alcohol.
● This wait time cannot be shortened by rinsing the mouth with
water or non-alcoholic drinks.
● Residual mouth alcohol or flavored drinks (e.g. fruit juice), breath
sprays containing alcohol, medicinal juices and drops can distort
the result.
● Belching and vomiting can distort the result.
● Drafts and climatic conditions such as wind, moisture and dust can
distort the result when performing measurements using the funnel.
● The seal must not be damaged.