9

Technical data

Measurement principle
Measuring range
Breath alcohol concen-
tration
Blood alcohol concen-
tration
A conversion factor of 2100:1 is used to calculate the blood alco-
hol concentration using the breath alcohol concentration data
(factory setting, country-specific conversion factors can be con-
figured).
Measuring units
mg/L, μg/100 mL, μg%, μg/L, g/210 L, g/L, ‰, %, mg/100 mL,
mg%, mg/mL
Ambient conditions
during operation
in storage
26
Electrochemical sensor
0.00 to 0.60 g/210 L (mass of alco-
hol per minimum volume of breath
at 34°C [93.2°F] and 1013 hPa)
0.00 to 0.60 % (mass of alcohol per
blood volume)
-10 to +55°C (14 to +131°F)
10 to 100 % r.h. (non-condensing)
600 to 1300 hPa
-20 to +60 °C (-4 to 140 °F)
15 to 75 % r.h. (10 to 100 % r.h. <48
h)