Puesta En Marcha - Vega VEGAPULS C 21 Instrucciones De Servicio

Sensor de radar para la medición continua de nivel
Ocultar thumbs Ver también para VEGAPULS C 21:
Tabla de contenido

Publicidad

Register Num-
Register Name
ber
203
Stopbits
206
Delay Time

14.4 Puesta en marcha

Register Num-
Register Name
ber
3000
Byte Oder (Floating point
format)
3200
Distance unit
3201
Temperature unit
3401
Min.-Adjustment
3403
Max.-Adjustment
3600
Medium type
3601
Application liquid
3602
Application solid
VEGAPULS C 21 • Protocolo Modbus y Levelmaster
Data type
Configurable Values
enum8
1 = None, 2 = Two
r/w
uint16
10 ... 250
r/w
Data type
Configurable Values
enum8
0=ABCD, 1=CDAB,
2=DCBA, 3=BADC
r/w
enum16
44=ft, 45=m, 47=in,
49=mm
r/w
enum16
32=°C, 33=°F, 35=K
r/w
float32
0 ... Measurement range
r/w
float32
0 ... Measurement range
r/w
enum16
0=Liquids, 1=Solids
r/w
enum16
0 =Storage tank, 1=
Storage tank (product agi-
r/w
tation), 2=Cargo tank,
3=Reactor tank, 4=Do-
sing tank, 5=Stilling pipe,
6=Bypass, 7=Outside
of plastic tank, 8=Outsi-
de of mobile plastic tank,
9=Level of open water,
10=Open flume, 11=Water
weir, 12=Demonstration,
13=Pump station, 14=Co-
llection tank
enum16
0=Silo, 1= Bunker Big,
2=Bunker fast filling,
r/w
3=Profil registration heap,
4=Breaker mill, 5=De-
monstration
14 Anexo
Unit
Default Va-
lue
1
ms
50
Unit
Default Va-
lue
0
-
m
-
°C
-
0
-
Measure-
ment range
-
Liquids
-
Storage
tank
-
Silo
53