NOTICE
Grinding the footshell
Premature wear resulting from damage to the footshell
► Do not grind the footshell.
► Inform the patient.
5.1 Applying/removing the footshell
Inserting the foot into the footshell
Recommended tools: 2C100 changing device for footshell
>
1) Pull the Spectra sock over the prosthetic foot, avoiding any wrinkles.
2) Slide the prosthetic foot into the footshell. Insert the carbon spring into
the slot on the toe insert (see fig. 3).
3) Push the prosthetic foot into the heel area of the footshell until it locks
into place (see fig. 5).
4) Place the connection cap on the footshell (see fig. 6).
Removing the footshell
Recommended tools: 2C100 changing device for footshell
>
1) Push the heel area of the footshell to the rear and pull the prosthetic foot
up.
2) Remove the prosthetic foot from the footshell.
6 Use
CAUTION
Use of the product without shoes
Falling due to slipping on slippery surfaces
► Do not use the product on slippery surfaces.
► Inform the patient.
6.1 Cleaning
CAUTION
Use of unsuitable cleaning agents or disinfectants
Impairment of functionality and damage due to incorrect cleaning agents or
disinfectants
► Only clean the product with the approved cleaning agents.
► Only disinfect the product with the approved disinfectants.
► Observe the instructions for cleaning and care.
► Inform the patient.
Ottobock | 13