Specifications

1-1 General Specifications

LCD display :
4000 counts, 3 3/4 digit LCD readout.
Measuring rate : 1.5 times / sec.
Overrange display :
"OL" is displayed for "Ω" functions,
Actual value is shown for "A" and "V"
function.
Automatic power off time :
Approximately 30 minutes after power on.
Low battery indicator :
< is displayed.
Power requirement :
9V battery type 6LR61 (IEC) Alkaline.
Battery life : ALKALINE 9V ; 200 hours
typical.

1-2 Environmental Conditions

Indoor Use only.
Calibration :
One year calibration cycle.
Operating temperature :
0°C ~ 30°C (≦80% RH)
30°C ~ 40°C (≦75% RH)
40°C ~ 50°C (≦45%RH)
Storage temperature :
-20 to +60°C, 0 to 80% RH (batteries not
fitted).
Overvoltage category :
IEC 61010-1 600V CAT.Ⅳ.
CAT.Ⅳ equipment of OVERVOLTAGE
CATRGORY Ⅳ is for use at the origin of
the installations. Note examples include
electricity meters and primary over-current
protection equipment.
Operating altitude : 2000m (6562 ft)
Conductor Size : 51mm diameter.
Pollution degree : 2
EMC : EN 61326-1
Shock vibration :
Sinusoidal vibration per MIL-T-28800E
(5 ~ 55 Hz, 3g maximum).