Warranty Rules - DIEL ME 100 V2 Manual De Instalacion Y Uso

Ocultar thumbs Ver también para ME 100 V2:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ENGLISH
The air cooling status is shoed as following:
• Led Air off: all cooling are off.
• Led Air on: one cooling system is active.
• Led Air blink: all cooling system are active.

WARRANTY RULES

The device has a warranty period time of 3 years from test date marked on the label
and at the end of this manual.
The warranty is valid only whether damages are due to manufacturing defects or to a
wrong calibration of probes.
We aren't liable for damages due to a wrong wiring of probes or to a wrong power
supply voltage (for example 400 Volt AC).
At any rate we aren't liable for damages due to the bad working of the equipment.
The reparations in guarantee, except different accord among the parts, will be
effected in our factory in MONTECCHIO MAGGIORE (VI).
ATTENTION
Don't effect dielectrical test or partial discharge on the electric machine with the device
inserted, avoid if possible to connect the device directly to the secondary of the
transformer to protect, otherwise the transient overvoltage can damage the device.
This is always clearer if the power supply voltage of the device is 230 VAC and in case
of condensers for the remaking.
TESTING CERTIFICATE
The testing procedure is effected as follows:
• Push-buttons working.
• Relays testing contacts
• Main mechanical checking
• Calibration and linearization at +30°C and +150°C.
• Testing for 48 hours with variable power voltage supply
• Control outpur 4/20 mA
• Insulating testing between earth and relays at 2,5 KV for 60"
-
-