Menu Control - Control Characteristics And Parameters; Dynamik Internal - Xylem SI Analytics CT 72/2 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7.4.
MENU CONTROL – Control characteristics and parameters
Menu level 1
Level 2

7.4.1. Dynamik internal

NORM
°C
temp. stability
APER
Setpoint
NORM
°C
temp. stability
APER
temperature ramp
The Thermostat is qualified for internal temperature control
Press the
parameters or values (prior to ENTER)
Parameter level
or
0.1 ... 99.9
3 ... 9999
0 ... 999
This parameter affects the temperature sequence only in case of
internal control.
Factory setting:
Possible parameters:
NORM
t
APER
With both settings constant temperature is achieved after
t
approximately the same time.
Menu functions
key if a parameter is to be retained. Correction function for
The parameter flashes, switch by pressing
and
The parameter flashes, set by pressing
+
The parameter flashes, set by pressing
+
The parameter flashes, set by pressing
+
APER (aperiodic)
Allows for reaching the setpoint faster – with setpoint
change or ramp function – but overshooting of up to 5 % is
possible.
Ramp function: the increase of temperature occurs
temporally offset and achieves the target temperature
without overshooting.
Setpoint change: The temperature increases at the same
rate, the target temperature is achieved without
overshooting.
85

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si analytics ct 72/4Si analytics ct 72/2 tt

Tabla de contenido