MENU ATC - Absolut Temperature Calibration

Example:
1-point calibration
°C
T 1
T
3-point calibration
°C
T 1
T
T
1 = Original curve
T
Menu-Ebene 1
Level 2
ATC serves to compensate a temperature difference that might occur between
circulator and a defined measuring point in the bath tank because of physical
properties.
Principle:
For ATC calibration, in steady state the bath temperature at the location of the
temperature sensor (CT) is determined at the respective adjusted working
temperature. This value is then set on the circulator in the menu >ATCalibration<
under menu item > CTEMP X <.
This can be a 1-point, 2-point or 3-point calibration.
M
°C
°C
M = Temperature measuring instrument with temperature sensor
B = Bath tank (INTernal)
T = circulator
CT = Temperature on measuring point
°C
TT = Temperature on Thermostat
Press the
parameters or values (prior to ENTER).
Parameter level
or
Menu functions
key if parameter is to be retained. Correction function for
The parameter flashes, switch by pressing
and
>NO< Carry out an ATC calibration
>YES< return to standard operation after
calibration.
88