List Of Commands - Xylem SI Analytics CT 72/2 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

10.3. List of commands

out-Befehle: Parameter bzw. Temperaturwerte einstellen.
Command
out_mode_01
out_mode_01
out_mode_01
out_mode_05
out_mode_05
out_mode_08
out_mode_08
out_sp_01
out_sp_02
out_sp_03
out_par_06
out_par_07
out_par_08
The out commands are valid only in remote control mode.
Command to set the working temperature > SETPNT 1< to 55.5 °C
Command to retrieve the working temperature > SETPNT 1< in_sp_00↵
Response from the circulator:
Parameter
Response of circulator
0
Use working temperature > SETPNT 1<
1
Use working temperature > SETPNT 2<
2
Use working temperature > SETPNT 3<
0
Stop the unit = R –OFF-.
1
Start the unit.
0
Set the control dynamics - aperiodic
1
Set the control dynamics - standard
xxx.x
Set working temperature. „SETPNT 1"
xxx.x
Set high temperature warning limit
xxx.x
Set low temperature warning limit
xxx
Xp control parameter of the internal controller. 0.1 ... 99.9
xxx
Tn control parameter of the internal controller. 3 ... 9999
xxx
Tv control parameter of the internal controller. 0 ... 999
Remote control
55.5↵
98
out_sp_00 ⇔ 55.5↵
„OverTmp"
„SubTmp"

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si analytics ct 72/4Si analytics ct 72/2 tt

Tabla de contenido