Troubleshooting
Malfunction
Possible cause
The charging cradle
The crawling plate is
LEDs indicate an
not positioned
error during the
correctly.
charging process or
do not light up at all.
No supply voltage
present
Camera is completely
Software malfunction Press the ON/OFF
non-functional (e. g.:
frozen image, camera
does not respond to
pressed keys).
48
Remedy
The crawling plate is
not sufficiently
fastened in place.
Retighten screw.
Check and correct
position of the
carrying strap.
Remove foreign
objects from under
the crawling plate or
in the charging cradle.
Place charging cradle
on level surface.
Check operation of
the power supply unit
and/or vehicle
connection cable.
button for 10 seconds
and switch camera
back on.