EN
Circulating Assistant
WARNING: DO NOT contaminate the handpiece.
The transfer shield guards the sterile handpiece from
contamination during power module installation.
1. Hold the power module by the handle.
2. Insert the power module through the transfer shield
into the open compartment.
3. Push the handle down.
6
WARNING: ALWAYS sterilize the contaminated
transfer shield before using it again.
4. Remove the contaminated transfer shield.
Sterile Assistant
WARNINGS:
▪ When closing the door, hold the door by the latch.
DO NOT touch the edges of the door or the internal
surface of the door that faces the power module.
▪ ALWAYS verify the door is securely closed to prevent
exposing the power module to the sterile field.
1. Holding the door by the latch, use the door to push the
power module handle down until it locks into position
against the power module.
7308-001-700 Rev-A