GreenLiner 2V / 4V
6
vertical laser line on the wall and measure the distance 3m
from point B to another point C. Point C must be on the verti-
cal laser line, it means the height of C point is 3m.
Measure the distance from point A to point C, if the distance
is over 2 mm, please, contact with seller to calibrate the laser.
Care and cleaning
Please handle measuring instrument with care. Clean with
soft cloth only after any use. If necessary damp cloth with
some water. If instrument is wet clean and dry it carefully.
Pack it up only if it is perfectly dry. Transport in original con-
tainer/case only.
Note: During transport On/Off compensator lock (3) must be
set to position "OFF". Disregard may lead to damage of com-
pensator.
Specific reasons for erroneous measuring results
•
Measurements through glass or plastic windows;
•
Dirty laser emitting window;
•
After instrument has been dropped or hit. Please
check the accuracy.
•
Large fluctuation of temperature: if instrument will
beused in cold areas after it has been stored in warm
areas (or the other way round) please wait some
minutes before carrying out measurements.
Electromagnetic acceptability (EMC)
•
It cannot be completely excluded that this instru-
ment will disturb other instruments (e.g. navigation
systems);
•
will be disturbed by other instruments (e.g. intensive
electromagnetic radiation nearby industrial facilities or
radio transmitters).
Laser class 2 warning label on the laser instrument
Laser classification
The instrument is a laser class 2 laser product accortding to
DIN IEC 60825-1:2007 and Class II in accordance with CFR
21 § 1040 (FDA). This tool may be used without need for
further protective measures.
7