It is possible, to restart the reading of cartridges by pressing on
user back to the cartridge loading step (7.5).
If the user presses on the key
If you wish to save data, press
If you wish to restart the test, press

7.7 Validation of the number of detected strips

If all the strips and their respective cartridges have been read correctly, the touchscreen will
display the number of detected strips (see screen hereunder).
Press
to go to the next step.
Press
if you wish to return the the strip loading step.

7.8 Validation of the protocol ID

The touchscreen displays the Protocol ID that has been detected on the barcodes.
If necessary, this value can be modified by using the up and down arrows (
Pressing of
will cancel the test and brings the user back to the strip loading step.
Press
to go to the next step. The touchscreen displays:
Close the cover of the machine, then confirm the closing by pressing
next step.
7.9

First washing step

A first washing step is necessary for the final preparation of the strips: they will be immersed
for 60 seconds into the Wash Buffer well of the cartridge. The touchscreen displays « Washing
step. Please wait » and the countdown of the 60 seconds.
Neptune Instrument
, the touchscreen displays "Invalid setup. Download data?"
.
.
19
. This action will bring the
or
).
and continue to the
Revision nr : 200710