The barcode of the
Reag's missing,
cartridge cannot be
reload?
read
Strips
The lot of the
expired/Reag's
strips/cartridges is
expired
expired
Wrong association of
Invalid Setup
strip lots and
cartridge lots
The machine has
Timeout
been stopped by the
restart ?
user dirng the
protocol
Other translations of this Instructions for Use can be downloaded from our website:
https://www.diasource-diagnostics.com/
Neptune Instrument
- Cartridge missing (the
user has not introduced
the right number on the
touchscreen or in the
cartridge holder)
- Label of cartridge not
correctly aligned
Use of a kit for which the
expiry date is not valid
anymore.
The used components are
not compatible.
The stop has been longer
than the authorized 30
seconds.
32
- Validate the reloading
and enter the correct
number of strips on the
touchscreen and in the
cartridge holder
- Take the cartridge out of
the holder and check the
label
Check the validity of the
components
Check the lot components
of the kit (see the CoA of
the kit)
It is compulsory to keep
the intervention under 30
seconds
Once this delay has
expired, the test is
automatically stopped and
has to be restarted by
clicking on « V »