The exposure time can also be modified by changing the sensitivity factor. Press keys
time , the actual sensitivity factor will be displayed.
Use keys
and
comes on. To quit this mode, press key
the teeth and both patient sizes.
The selected operating mode is always used for each tooth and patient size combination in modes En60, En63 and
En65.
In AUTO mode, each tooth and patient size combination is associated to the best mode from amongst the ones
available. In this mode it is not possible to assign a mode other than the default one to each combination.
To set the mode, refer to paragraph 4.5 "Setting the mode and exposure time in USER mode".
To change the mode amongst En60, En63, En65 and AUTO refer to paragraph 5.2 "Setting the operating mode".

4.5. SETTING THE MODE AND EXPOSURE TIME IN USER MODE

In USER mode, it is possible to assign an exposure time and a mode from amongst En60, En63 and En65 to each
tooth-patient size combination.
The default setting corresponds to the AUTO mode settings with sensitivity factor F=19.
To activate USER mode regardless of the mode currently being used, press keys
Icon
will come on to signal USER mode is active.
To deactivate USER mode press keys
The exposure parameters are set by following the directions given below::
1) Select the tooth to be examined
2) Select the patient size
The exposure time is automatically shown on the handheld screen.
NOTE: It is not possible to access the sensitivity factor menu in USER mode. In addition, keys
are inoperative in User mode.
The exposure times and mode assigned to the tooth – patient size combinations are custom set by following the
directions given below:
1) Press and hold down key
2) Select the desired tooth-patient size combination.
3) The exposure time can be changed with keys
NOTE: Exposure times ranging from 0.01s and 1.00s belonging to the R'20 scale can be set.
70
to change the value from 3 to 25. If the displayed value differs from the default setting, icon
or
and
about two seconds. Customized settings can be entered and icon
OPERATOR'S MANUAL
. The change made to the sensitivity factor, is applied to all
again (icon
goes off).
and
.
and
at the same
and
at the same time.
comes on.