4.
With the power switch in the "Off" position, disconnect the device from all external power sources.
5.
Visually inspect the device for any damage, missing parts, etc.
6.
Clean the exterior of the device with a clean lint-free cloth. Heavy soil should be removed with a clean lint-free cloth dampened with water. A soft bristled brush dampened with
water can be used to remove stubborn soil. Dry the device using a clean lint-free cloth if water was used to remove soil.
7.
Use 5.25% chlorine bleach (Clorox Regular Liquid Bleach or equivalent). Mix one (1) part bleach with four (4) parts water in an appropriate clean container. This ratio produces a
one (1) part bleach to five (5) total parts solution (1:5). The total volume (amount) of solution required is determined by the number of devices in need of disinfection.
An alternate suitable commercial (bacterial-germicidal) disinfectant that meets the requirements listed in the NOTE below and is used as per the disinfectant manufacturer's
recommended dilution rates and instructions may also be used.
CAUTION
Do not submerge the device in water as this will result in damage to the vacuum pump.
NOTE– Do not use any cleaners or disinfectants that contain ammonia, benzene and/or acetone to clean the device.
8.
Apply the bleach solution in an even manner to the enclosure and power cord using a clean lint-free cloth. The cloth should be dampened only and not dripping of solution. Do not
use a spray bottle to apply the solution. Do not saturate the device with the solution. Avoid over-saturating the enclosure seams so that no solution residue builds up in these
areas.
9.
Exposure time of the disinfectant solution should be 10 minutes minimum to 15 minutes maximum.
10. After the recommended exposure time, all exterior surfaces of the device are to be wiped with a clean lint-free cloth dampened with drinking quality water no warmer than room
temperature. Dry the device with a dry, clean lint-free cloth. This is to remove residue that may stain or leave a film on the device, especially after repeated disinfections.
Disinfection for Multi-Patient Use
NOTE– The disinfection process can only be completed by the manufacturer or by an appropriately trained individual.
Enclosure, power cord
Collection container, filter, tubing and carrying
case
EN - 16
Recommended disinfection interval
Between patients
Do not clean, replace between patients
Number of disinfection cycles
30
N/A
Compatible disinfection method
1:5 chlorine bleach (5.25%) and water solution
N/A
NOTE–
A-7325