Battery Warnings - Drive DeVilbiss Healthcare Vacu-Aide 7325 Serie Guía De Instrucciones

Unidad de succión
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

WARNING
1.
Improper use of the power cord, adapters, and plugs can cause a burn, fire, or other electric shock hazards. Do not use the unit if the power cord or adapters are damaged.
2.
Locate tubing and power supply cords to prevent tripping hazards and reduce the possibility of entanglement or strangulation.
3.
When device is used under extreme operating conditions, the temperature near the exhaust vents on the bottom of the unit may reach 70˚C (158˚F). Keep bare skin away from
this area.
4. The following surface temperatures may exceed 41 ˚C under extreme conditions:
• Collection Container ............................................................................. 42.7 ˚C (108.9 ˚F)
• AC/DC Power Supply ........................................................................... 43.3 ˚C (109.9 ˚F)
• External Surface of Suction Unit (Inside of carrying case) .................. 62.2 ˚C (144.0 ˚F)
• Power Switch.........................................................................................47.8 ˚C (118.0 ˚F)
• Bottom Surface of Handle .....................................................................48.8 ˚C (119.8 ˚F)
• Top Surface of Underneath Handle ..................................................... 52.4 ˚C (126.3 ˚F)
5.
To avoid electric shock, do not remove the cover. The cover should only be removed by a qualified Drive DeVilbiss technician.
6.
Before attempting any cleaning procedures, turn the unit "Off" and disconnect from AC or DC power.
7.
Do not service or clean this device while in use with a Patient.
WARNING
This suction equipment should only be used by persons who have received adequate instructions in its use.

BATTERY WARNINGS

1.
Charge the battery before initial use.
2.
Recommended maximum time between charges = 1 year
3.
Recommendation: Store the battery below 60˚C (140˚F), low humidity, no dust and no corrosive gas atmosphere. Store fully charged if possible.
4.
This device contains electrical and/or electronic equipment. Follow local governing ordinances and recycling plans regarding disposal of device components.
5.
The battery must be recycled or disposed of properly.
CAUTION
Should fluid be aspirated back into the unit, equipment provider servicing is necessary as possible vacuum pump damage may result.
NOTE– This device and accessories, as packaged, do not contain natural latex rubber.
NOTE– Always verify suction level before beginning by occluding open end of patient tubing while observing gauge. Adjust knob to desired level.
NOTE– To minimize the ENVIRONMENTAL IMPACT of the 7325 Series Suction Device during NORMAL USE, remove tubing and collection container from the suction unit, clean or
dispose the tubing and collection container properly. Unplug the 7325 Series Suction Device from AC power. Clean the cabinet as listed in the Cleaning section of this guide.