Using the Monitor
Transmitting Temporarily Stored Data
In cases when the receiver device cannot receive measurement
data, the measurement data is temporarily stored in the monitor
memory. The data stored in the memory is transmitted the next time
a connection is successfully made to the receiver device. A total of 30
sets of measurement data can be stored. When the amount of data
exceeds 30 sets, the oldest data is deleted and the new data is stored.
Time
This monitor has a built-in clock. The date and time a measurement was
taken is included in the measurement data. The built-in clock is designed
to be automatically adjusted by syncing with the clock of a receiver
device. Refer to the specifications of the receiver device. This monitor
has no clock adjustment function.
Selecting the correct cuff
Using the correct cuff size is important for an accurate reading. If the
cuff is not the proper size, the reading may yield an incorrect blood
pressure value.
The arm size is printed on each cuff.
The index p and proper fit range, on the cuff, tell you if you are applying
the correct cuff. Refer to "Applying The Arm Cuff".
If the index p points outside of the range, contact your local dealer to
purchase a replacement cuff.
The arm cuff is a consumable. If it becomes worn, purchase a new one.
Part Number
UA-420A
UA-291A
UA-290A
UA-289A
Arm size: The circumference at the biceps.
Description
Wide Range cuff
Large cuff
Medium cuff
Small cuff
Specification
8.6-16.5" (22 to 42cm)
12.2-17.7" (31 to 45cm)
9.0-14.6" (23 to 37cm)
6.3-9.4" (16-24 cm)
E-14